07/02/01 1-1 DIGIBAR-Pro PROFILING SOUND VELOCIMETER OPERATION MANUAL Software Version 1.XX
07/02/01 2-92.3 Initial Configuration Power ON Depress and hold the power button for 3 seconds. The display will show the Header Screen for 1
07/02/01 2-102.4 Setup Menu Parameters Log Basis: Velocities can be logged in the control unit based on either depth or time. The usual method of p
07/02/01 2-11influence of single bad data points, or storing one data point at each predetermined measurement interval. The results of a cast using a
07/02/01 2-122.5 Performing a Velocity Cast From the Main Menu, position the arrowhead cursor opposite the Acquire parameter and press the Enter key
07/02/01 2-13Once the velocity/depth probe has been lowered to the designated depth and then retrieved, position the cursor opposite the Cast paramete
07/02/01 2-14 Pressing the Enter key a second time will delete the record from the cast. . 2.6 Acquire Parameters In the Acquire mode, several
07/02/01 2-15rinsed with fresh water and dried after each cast. This will minimize the possibility of corrosion affecting the function of the probe.
07/02/01 3-163 Interfacing 3.1 Uploading Information to a PC Main Menu After completing a Cast, it will often be desirable to upload the r
07/02/01 3-17Output Type: The choice of 5 different output formats can be made using the this parameter. CSV: Selecting the CSV format option results
07/02/01 3-18 WESTGEO: This sentence complies with the requirements of the current WG-1100 velocimeter output string. 634 01503 635 01503 636 01504
07/02/01 1-2TABLE OF CONTENTS 1 INTRODUCTION ... 1-
07/02/01 3-19DIGIBAR-PRO control unit and an external computer. Selection of Yes under Output During will cause this to happen. The output string w
07/02/01 4-20chart, and press the Enter key. This will cause the Echotrac/Hydrotrac to print the information contained in the Sounder string on the u
DigibarPro Version 1.07 Appendix1 10/27/08 4-21Appendix 1 Downloading Examples using Digibar Pro Log 2.1: HYPACK compatible velocity tables: Dow
DigibarPro Version 1.07 Appendix1 10/27/08 4-22b. The box next to Com Port is checked and no checks remain in the boxes next to the "direct
DigibarPro Version 1.07 Appendix1 10/27/08 4-23 8. Click on the box next to the Begin Cast Download button and enter the # of the particular ca
DigibarPro Version 1.07 Appendix1 10/27/08 4-2410. Click on the Graph tab. Observe that the graph of the downloaded data looks appropriate for t
DigibarPro Version 1.07 Appendix1 10/27/08 4-25 An interesting observation: In Hypack applications, if velocity data is collected in feet/sec.,
DigibarPro Version 1.07 Appendix1 10/27/08 4-26Once this step is completed, additional editing or plotting of the sound velocity profile can be a
DigibarPro Version 1.07 Appendix1 10/27/08 4-27Appendix 2 Quality Assurance Check To receive the utmost reliability and accuracy from your Digib
DigibarPro Version 1.07 Appendix1 10/27/08 4-28
07/02/01 1-2LIST OF FIGURES Figure 1. 1 ...
07/02/01 1-3 1 INTRODUCTION The Operation manual contains information required to operate and field test the DIGIBAR-PRO Model 1200 hand-held sound v
07/02/01 1-4The velocity of sound and depth values displayed on the DIBIGAR-PRO front panel are measured, stored in memory, and displayed by the syste
07/02/01 1-51.3 Features The following is a list of standard DIGIBAR features: • Velocity and depth displayed in feet or meters • Automatic stora
07/02/01 1-61.4 Control Unit Front Panel Description All controls on the unit’s front panel as well as the LCD display are sealed to prevent water d
07/02/01 1-71.5 Probe/Cable Description The velocimeter/depth probe, in its standard configuration, is connected to the hand-held control unit via a
07/02/01 2-82 OPERATION Both the membrane switches and display on the control unit's front panel are sealed, and all openings in the body of the